Dose Distribution in Low Energy Exposure and Measurement Accuracy with Passive Dosimeters used in Individual Radiation Protection Dosimetry
F. Mihai, A. Stochioiu
Horia Hulubei, National Institute for R&D in Physics and Nuclear Engineering, Magurele, Ilfov
Full Text PDF
The work proposes the study on the radiation distribution uniformity provided by a 241Am standard source. The Agfa films and LiF:Mg,Cu,P thermoluminescent dosimeters were irradiated at doses from 0.01รท10.0 mSv range. All detectors were placed on the PMMA phantom within of the radiation solid angle. In point of 1.00 mSv conventional dose the calculated dose was from 1.07 mSv to 0.64 mSv from the middle towards the edge of the solid angle. There are differences of doses into the solid ange of radiation.

DOI:10.12693/APhysPolA.135.1081
topics: dose, measurement accuracy, dose distribution, optical density